Cyclic Compression Porometer

BACK
Share

Standards:

Applications:

The Cyclic Compression Porometer is used to test the sample and measure the maximum pore diameter, mean flow pore diameter and pore distribution. Theses data represents the actual characteristics of pore structure.
Applied industries:
Automobile manufacturing      Fuel battery pack       Paper     Battery separator     Geofabrics
Filtration            Nonwoven              Wool felt        
 
 

Product Information:

Introduction

The instrument applies cyclic stress on a sample and measures the pore structure characteristics after a desired number of cycles. The stress cycles and testing after any desired number of cycles are automatically continued.

Operation Principle
The instrument applies cyclic stress on a sample and measures the pore structure characteristics after a desired number of cycles. The sample is loaded in the sample chamber and is subjected to stress cycles in the specified stress limits. At the end of the desired number of stress cycles, the pressure of a non-reacting gas on one side of the sample is increased to initiate gas flow through pores. The gas pressure and flow rates are measured. The pores in the sample are spontaneously filled by a wetting liquid. The gas pressure and flow rates are measured through the wet sample. After acquisition of data, the sample is re-wetted and again subjected to cyclic compression. Pressure and flow rates are measured after the desired number of cycles. The test is continued to acquire data as a function of number of stress cycle.
 
 
 
 Features
 · After desired number of stress cycles: automatically interrupts analysis, performs tests, acquires data, continues to cyclically stress the sample
· Very little operator involvement
· Operator adjustable: stress limits in a cycle; number of cycles after which data is to be acquired ; stress free duration for data acquisition ; total number of cycles.
· Concurrent measurement of compressive stress in the sample as a function of stress cycle.
· Windows based simple operation.
 
Technical Parameters
Items Parameters
Pore Size Range 0.013 - 500 microns
Permeability Range 1 x 10-3 - 50 darcies
Sample size 1.75” - 2.5” diameter
Pressure range 0~500psi
Pressurizing gas Clean, dry, and compressed air or nonflammable and noncorrosive gas
Pressure transducer range 0~500psi
Resolution 1 in 60,000
Accuracy 0.15% of reading
Mass flow controller range 10cc/minute~100cc/minute
Power supply 110/220VAC,50/60Hz (Others Available)
Dimension D30” H x 19” W x 18.5” D
Weight 100lbs
 