Oil Mist Separator Test System

BACK
Share

Standards:

Applications:

Product Information:

The test rigs purpose is the testing of oil mist separators re capture rate, flowrate curve (pressure loss relative to flowrate), pressure regulation curve and leakage. The unit to be tested will be fed with an oil aerosol (flowrate and temperature adjustable) either in positive or negative pressure conditions. The immediate vicinity of the filter is temperature controlled. During the collection efficiency measurement the separated and the let-to-pass oil amounts are measured in dependence on the volumetric flow. They are automatically recorded, applying the gravimetric method, by means of an absolute filter. The amount of oil leaving the separator can be recorded online. Additionally, the leakage of the oil drain valve can be measured. The test rig simulates the operating conditions in typical fields of oil aerosol separator application. When recording the pressure control characteristic, the pressure in the crankcase“ is simulated in dependence on the „suction pipe pressure“ for different flow rates.
 
Features:
 ◆ Wide range of flow rate for overpressure and underpressure operation mode
◆ Special operating mode for testing pressure control valves
◆ Sample holder adjustable for different sample dimension
◆ Heatable test chamber (ambient temperature up to 120°C/248°F)
◆ Closed loop test air recovery system (with cooling stage)
◆ Aerosol generation by means of ATM 243 and/or LDG 244
◆ Safe test chamber with hinged cover
◆ Software guided testing procedure
◆ Convenient Windows software for test rig control, data acquisition and data export

Technical Parameters:
Items Parameters
Flow rate 15...300 l/min
Alternative 20…400 l/min
Pressure range of unit under test ± 150 mbar
Differential Pressure 200 mbar
Testing liquid Mineral or synthetic engine oil
ATM 243 Generated oil  max. 5...60 g/h  Aerosol d50.3 approx. 1.3 μm  Nozzle temperature (176°F … 248°F)
Option LDG 244 Generated oil 50 … 2000 g/h  Aerosol d50.3=50…100 μm Nozzle temperature 90°C .... 120°C
Temperature range of the testing chamber Ambient air temperature up to 120°C (248°F)
Supply air quality requirements Room air (common laboratory grade)
Cooling water Tap water; max. 6 bar, max. 20 l/h, cooling circuit (outlet temperature max. 40°C)
Compressed air supply Max. 6 bar; 300 l/min (alternative 400 l/min)
Power supply 400 V, 16 A
Dimensions (L x W x H) 2100 x 925 x 1200 mm
Weight Approx. 400 kg