Search keywords:

product name, product type, model number,
test method, manufacturer, technique, application

Automated Filter Tester G210

BACK
Share

Standards:

US 42 CFR part 84,EN 143,JMOL,ISO 23328-1

Applications:

Automated filter testing
 

Product Information:

Automated Filter Tester Model G210 meets NIOSH’s proposed Powered Air Purifying Respirator (PAPR) filter test requirements. It is an oil-only tester that has an airflow range of 100 to 200 L/min and can measure efficiencies up to 99.999%, or penetrations as low as 0.001%. In fact, Model G210 uses the same proven oil-aerosol generator found in models 27 and 8130. Like models 27 and 30, Model G210 features dual light scattering photometers that simultaneously measure upstream and downstream particle concentration to ensure fast, repeatable results.
 
Features:
◆ A touch-panel display makes the Model G210 easy to use, so very little operator training is needed. On-screen menus enable users to easily change test parameters to meet specific requirements of different filters. A microprocessor controls the tester and calculates and displays filter penetration, resistance value and flow rate. Self-check diagnostics and automatic zeroing between tests ensures accurate, stable, and reproducible results.
◆ Compliance with respirator regulations
◆ Simple, fast and automated operation
◆ Service and support
 
Components:
1. Dust removal system:
The compressed filtration air system provides a certain amount of testing dust for the aerosol generator
2. The test chamber:
Main test chamber and sample clamp
3. Monitoring and regulating system:
Pressure monitoring and adjustment provide dynamic test environment
4. Results display and analysis system:
Monitor through aerosol and display test results
 
Technical Parameters:
Items Parameters
Measurement Application Loading and Quality Control Tests
Maximum Efficiency 99.999%
Aerosol Type DOP/PAO/DEHS/Paraffin/other Oils
Aerosol Generation Atomizer
Count Median Diameter 0.2μmm (Oil)
Geometric Standard Deviation <1.6m(Oil)
Flow Rate 15-100L/min
Resistance 0-150mm H2O(0-1470Pa)
Particle Detection Light Scattering Photometer
Typical Test Length 10s
Data Reporting Alpha-Numeric Display, RS-232
Operation Stand Alone Tester/Automated Production Lines
Compliance US 42 CFR part 84,EN 143,JMOL,ISO 23328-1

Leave Message