Search keywords:

product name, product type, model number,
test method, manufacturer, technique, application

Automated Filter Tester 30

BACK
Share

Standards:

USA commercial respirator regulation (42 CFR part 84) European EN 143, ISO 16900-3 and related respirator standards? View EN 143 Compliance Report (US) View EN 143 Compliance Report (DE) ISO 23328-1 (testing filters for anaesthetic use) JMOL (Japanese respirator standard)

Applications:

Product Information:

The Automated Filter Tester Model 30A continues to be the best solution for testing particulate respirator filters, disposable filtering face pieces, and a wide assortment of filter media. The filter testers have been used in quality control and manufacturing by leading filter and filter media manufacturers and test centers around the world for more than 20 years.
Based on established model 30, the world's most-used filter tester, model 30A now offers higher sensitivity and resolution - up to 99.9999 percent efficiency has been produced with Emery oil at concentrations of 200 mg/m³ - at a lower cost of ownership due to the new serviceable photometers inside. The model 30A features a high degree of automation and self-diagnostics that greatly simplifies operation, increases throughput, and improves overall measurement performance. Whether looking for a stand-alone tester to check a few filters per day or need to integrate a tester in an automated, high-volume production line, the model 8130A is up to the challenge.

Features and Benefits:
  • Complies to more standards than any other filter tester
  • Salt and oil testing in one unit 
  • Highly consistent test results
  • New user-interface and data recording options
  • Customizable fully automated testing sequences
  • Ergonomic design
  • Reduced cost of ownership with new serviceable photometers
  • Worldwide service and support
Technical Parameters:
Items Parameter
Application Load and quality control
Max Efficiency 99.999%
Aerosol Type DOP/PAO/DEHS/Paraffi/Others Oil Aerosol、Salt
Aerosol Generation Atomizing
CMD 0.2μm(Oil)or 0.75um(NaCl)
GSD <1.6(Oil)or <1.86(NaCl)
Flow Rate 15-100L/min
Resistance 0-150mm H2O(0-1470Pa)
Particle Detect Scatter Photometer
Measurement Time 10 Sec.
Data Report LCD,RS-232
Operation Independent tewster、automatic production line
Standard compliant US 42 CFR part 84,EN 143,JMOL,ISO 23328-1 ISO 16900-3
Power Supply 115VAC、60Hz,10A或230VAC,50Hz,5A
Compressor Air 80psig,7scfm(550kPa,198std.L/min)
Dimension 70cm x 70cm x 147cm(28in, x 28in.x 58in.)
 