Humidified Tandem Differential Mobility Analyzer

BACK
Share

Standards:

Applications:

· Study of the hygroscopic growth of aerosols for accurate deliquescence and efflorescence measurements. 
· Other aerosol studies by the TDMA method.
 

Product Information:

Introduction

The Humidified Tandem Differential Mobility Analyzer (DMA) system has two differential mobility analyzers and a humidification section in between. Aerosol from the ambient atmosphere or from the built-in aerosol generator can be classified by the first DMA to produce a monodisperse aerosol for size analysis by the second DMA to determine the growth in particle size as a function of relative humidity.
 
Structure
· Built-in Atomizer Aerosol Generator
· Electrical Ionizer
· Differential Mobility Analyzer (DMA)
· Electrical Mobility Spectrometer
· Condensation Particle Counter (CPC)
· Humidity Conditioner
· Data analysis software

Features
· High stability atomizer aerosol generator
· High accuracy aerosol humidity conditioner
· Wide range, high resolution DMAs
· Patented dual-reservoir CPCs
· DMS and SMS data analysis software
· User friendly user interface with touch screen display
· Automatic data logging

 Technical Parameters
 
Overall Specifications of the Model 1040XP HTDMA*
Aerosol Generator (Part 1 of system) 460 mm (H) x 610 mm (D) x 525 mm (W)
Part 1/Part 2 Total Weight 100 lbs (45 kg)/ 140 lbs (60 kg)
Power Requirements (Part 1) 115 or 230 VAC, 50-60 Hz, single phase, 250 W max, 160 W steady
Humidity Conditioner/Spectrometer (Part 2) 460 mm (H) x 610 mm (D) x 525 mm (W)
Power Requirements (Part 2) 100- 240 VAC, 50-60 Hz, single phase, 500 W max, 400 W steady
Operating Aerosol Pressure 800 to 1050 mbar absolute pressure
Ambient Temperature Range 10 to 35 ºC
Ambient Humidity Range 0-90% RH, non-condensing
Polydisperse Aerosol Generation System
Single-Jet Atomizer Stainless Steel Construction
Pressure 5 to 60 psig (5 to 415 kPa gage), controlled by manual pressure
regulator
Aerosol Flow Rate Typical 3 L/min @ 35 psig (240 kPa gage)
Volume Mean Droplet Diameter about 2 µm (GSD< 2.0)
Aerosol Size Range 10 to 2000 nm
Aerosol Concentration > 107 particles/cm3 (typical for NaCl)
Dryer Diffusion type (NafionR membrane requires dry comp. air, -20 ºC
dew point)
Humidity Conditioner
Relative Humidity Control Range 10 to 93% with user-selected set point in 1% intervals
Flow Rate 1 to 10 L/min
RTD and Dew Point Accuracy ± 0. 2 ºC
RH Accuracy ± 1.5% for chilled mirror sensor; ± 2.5% for capacitive sensor
RH control stability ± 1.2% (typical)
Response Time for < 10% ΔRH < 3 min (typical)
Water Reservoir Heater Control 25 to 50 ºC in 5 ºC intervals
Saturator Nafion® Humidifier with DI water closed loop circulation
Dryer Diffusion type (Drierite® desiccant)
Differential Mobility Analyzer
Inlet Aerosol Flow Rate 0.1 to 1.0 L/min, controlled externally from Model 1500 AGM
Monodisperse Aerosol Flow Rate 0.1 to 1.0 L/min (equal to inlet aerosol flow rate)
Sheath Air and Excess Air 1.0 to 9.0 L/min, closed-loop, user-selected set point
High Voltage Power Supply 10 to 10,000 V, user-selected (software limits max voltage to 9,000
V)
Aerosol Size Range 8 to 1000 nm, user-selected
Maximum aerosol input conc . 108 particles/cm3
Particle Type Solids and non-volatile liquids
Condensation Particle Counter
Working Principle Laminar continuous flow with single particle detection
Flow Rate 0.3 L/min controlled by internal feedback control system
Particle Diameter Range 5 nm to 2 µm
Concentration Accuracy ± 10% (compared to calibrated reference MSP CPC)
Concentration Range 0 to 3x104 particles/cm3; (single particle counting with coincidence
correction)
Working Fluid 1-butanol (n-butyl alcohol)
Response Time (95% response) < 5 sec
*Specifications are subject to change without notice .
 