The medical mask flame retardant tester is mainly used to test the combustion performance of medical masks when they come into contact with flames at a certain linear speed. It is a medical mask flame retardant performance testing instrument.
Compliant with standards:
GB19083-2010 "Technical Requirements for Medical Protective Masks", YY0469-2011 "Medical Surgical Masks", GB2626-2006 "Respiratory Protective Equipment Self suction Filter Particle respirator" 6.15 Flammability, BSEN136-1998, GB19083-2003 "Technical Requirements for Medical Protective Masks - Flame retardant Performance"
Technical parameters:
1. The automatic motion line speed of the head mold is (60 ± 5) mm/s;
2. The flame height can be adjusted by 40 ± 4mm;
3. Equipped with flame temperature measurement and detection thermocouples;
4. The temperature of the flame at a height of (20 ± 2) mm from the top of the burner is (800 ± 50) ℃;
5. The distance between the top of the burner and the low point of the mask is (20 ± 2) mm;
6. The height adjustment range of the burner is 30mm;
7. The wall thickness of the metal head mold is 2-3mm, which is flame resistant and combustion resistant;
8. Continuation and smoldering time timers: 0-9999.9 seconds, accuracy: ± 0.1 seconds;
9. Working power supply: AC220V, 50Hz, 50W.