Product Introduction:

The purpose of this instrument is to measure the temperature change of the sample with radiant heat.

Arrange the sample and its reference material on the sample table, install it on the measuring table at an Angle of 45°, set a heating plate parallel to the measuring table, measure the surface temperature of the two samples with a radiation thermometer, etc., and measure by regression extrapolation method.

 

Specification parameters:

Model: IT-PT

Sample size: 10 X 7cm

Compare sample size: 10 X 7cm

Heater: Aluminum black plate

Heater operating temperature: 90℃

Heater control mode: PID control mode

Measuring table Angle: 45 degrees (black painting)

Sample station/heater spacing: 15cm

Equipment calibration: Measuring table (including sample mounting rack and sample table)

Heater (including temperature controller)

Heater storage box

Measuring chamber conditions: 20±1℃, 50± 10%RH

Power supply: AC100V or 220v (using downlink transformer), 50hz or 60hz, single phase

Attention: Thermal imaging {detection wavelength range 8 —— 13μm/ detector Hg, Cd, Te system}, controller {temperature resolution 0.05℃, emissivity 1.0, scanning frame time 0.6 —— 0.8 seconds, image processing (setting of start time, display of average temperature inside the box)}, monitor {high resolution monitor or LCD color display with built-in controller}, please prepare yourself.

 